HD-E808-160A Salt Spray Chamber
EQUIPMENT DETAILS OF SALT SPRAY CHAMBER
Inner chamber
GENERL SPECIFICATIONS
Packaging Dimensions: (WxDxH) 2550*1550*1650mm
Power supply source: single-phase, 220V±10%, 50Hz (can be appointed)
Gross Weight: 300kg
HS Code: 8479899990
STANDARD PARAMETERS
Model | HD-E808-160A |
Internal Size (W*D*H) | 1600*1000*600mm |
External Size (W*D*H) | 2300*1300*1400mm |
Equipment Material | P.V.C rigid plastic plate, thickness 8mm |
Sample tray | Diameter 10mm glass fiber rod, anti-rust V-shape plastic steel to make sure sample in 15~30 degree inclined |
Temperature Range | Room temperature ~ 55℃ |
Temperature Fluctuations | ≤±0.5℃ |
Temperature Uniformity | ≤±2℃ |
Temperature Precision | ±1℃ |
Spraying method | Continuous and Periodic spray type |
Test Chamber Temperature | Salt Spray Method (NSS ACSS)35±1℃ |
Saturated air Barrel Temperature | Salt Spray Method (NSS ACSS)47±1℃ |
Brine Temperature | 35℃±1℃ |
Spray Quantity | 1.0~2.0 ml / 80cm2 / hr |
PH value | Salt Spray Method (NSS ACSS6.5~7.2) |
Lab Volume | 800L |
Brine Tank Capacity | 40L |
Multiple Safety Protection Devices | Current discharge protection, over pressure protection, over temperature protection, over load fuse protection |
Accessories | Testing Salt / Measuring tank / Spray Nozzle x 1 set |
Air source | 1HP Air pump (provided by buyer) |
Test criteria | ASTM B117-08, ISO 9227... |
PRODUCT FEATURES
1. Durable hard 5 mm PVC construction, anti-corrosion and easy to clean.
2. Reservoirs with mixing system on stand.
3. Separate heating system apart from salt testing chamber to ensure accurate result and protect circuit.
4. Exposure zone lined with corrosion-resistant PVC sheets.
5. Clear cover allows viewing of test conditions.
6. Peaked cover prevents dripping of condensation onto samples.
7. Water seal and water level protection system.
8. Heated humidifying tower with air relief valve, adjustable tower tip to orient salt fog.
9. Automatic DI water fills to humidifying tower.
10. Digital temperature control.
11. Integral over-protection on heaters.
12. Low water cut-off protection on humidifying tower.
EQUIPMENT IMAGES
Sodium chloride (testing salt) Measuring tank
Spray nozzle
EQUIPMENT DETAILS OF CLIMATE CHAMBER
Temp. Humid. Programmable Controller
German Bock compressor
GENERL SPECIFICATIONS
Packaging Dimensions: (WxDxH) 2400*1680*1900mm
Power supply source: three-phase, 380V±10%, 50Hz (can be appointed)
HS Code: 8479899990
STANDARD FEATURES
Model | HD-E702-1000B70 |
Internal dimension (W*D*H) | 1600*1100*800mm |
External dimension (W*D*H) | 2100*1480*1700mm |
Temperature range | 0~+30℃ |
Temperature fluctuation | ±0.5℃ |
Temperature uniformity | 2℃ |
Humidity range | 20~60%RH (Refer Image below) |
Humidity fluctuation | ±2.5%RH |
Humidity uniformity | 3%RH |
Cooling speed | 1℃/min in average (without loading) |
Heating speed | 3℃/min in average (without loading) |
Internal chamber material | SUS#304 stainless steel, mirror finished |
External chamber material | Stainless steel |
Cooling method | Air cooling |
Controller | LCD touch screen, programmable control temperature and humidity Can set different parameter for cyclic test |
Insulation material | 50mm high density rigid Polyurethane foam |
Heater | Explosion-proof type SUS#304 stainless steel fins radiator pipe heater |
Compressor | France Tecumseh compressor x 2sets |
Lighting | Heat resistance |
Temperature sensor | PT-100 dry and wet bulb sensor |
Observation window | Tempered glass |
Testing hole | Diameter 50mm, for cable routing |
SUS#304 stainless steel, 2pcs | |
Safety protection device | Protection for leakage Over-temperature Compressor overvoltage and overload Heater short circuit Water shortage |
APPLICATIONS
1. To simulate test environment with different temperature and humidity
2. Cyclic test includes climatic conditions: holding test, cooling-off test, heating-up test, moistening test and drying test…
3. Cable port with flexible silicone plug for cable routing to provide the condition of test unit under operation
4. Uncover test unit weakness in a short-term test with accelerated time effect
CHAMBER DESIGN FEATURES
1. High performance and quiet operation (68 dBA)
2. Space-saving designed for flush installation to wall
3. Full thermal break around doorframe
4. One 50mm diameter cable port o left, with flexible silicone plug
5. Accurate wet/dry-bulb humidity measurement system for easy maintain
PROGRAMMABLE CONTROLLER
1. PLC controller for test chamber
2. Step types include: ramp, soak, jump, auto-start, and end
3. RS-232 interface to connect computer for output
PLC OPERATION
INTERNAL CHAMBER