Product Description
This calibration device is designed in accordance with ASTM D5207-03 and related IEC, UL, GB standards for measuring flame temperatures in combustion test equipment.
Calibration Procedure:
Position the copper block 55±1mm above the needle flame burner.
Ignite the flame; the time for the copper block temperature to rise from 100°C±2°C to 700°C±3°C must be 54±2 seconds to validate the flame for testing.
Perform three measurements, allowing the copper block to cool naturally to <50°C between tests.
Reference Standards
ASTM D5207-03: Standard Practice for Confirmation of 20-mm (50-W) and 125-mm (500-W) Test Flames for Small-Scale Burning Tests on Plastic Materials.
GB/T 5169.15: Fire Hazard Testing for Electric and Electronic Products — Part 15: Test Flames — 500W Flame Apparatus and Confirmational Test Methods.
IEC 60695-11-3: Fire hazard testing for electric and electronic products — Part 11-3: Test flames — 500W flames — Apparatus and confirmational test methods.
Product Specifications
| Model | 6015-HD |
| Thermocouple | Armored K-type (NiCr/NiAl) fine-wire thermocouple with a 0.5mm diameter sheath. |
| Timer | Adjustable range: 0.01–99.99 seconds |
| Copper Block Specifications | |
| Diameter | Ø9.00±0.01mm |
| Length | 19.14±0.02mm |
| Bevel Angle | 45±0.5° |
| Mass (Pre-drilling) | 10.00g±0.05g |
| Material | High-conductivity electrolytic copper |
| Surface Finish | Fully polished |