Needle Flame Test Instrument

  1. Inicio
  2. /
  3. Equipo de ensayo de inflamabilidad
  4. /Needle Flame Test Instrument

Needle Flame Test Instrument

Descripción

Descripción del producto
This instrument simulates small flames generated under fault conditions (e.g., leakage currents, overheating, poor contacts) to assess fire hazards.
It is applied when glow-wire or horizontal-vertical flame tests are unsuitable for small components. If a sample ignites during glow-wire testing, the needle flame test further evaluates combustion risks. Suitable for parts of electrical/electronic products and household appliances (e.g., terminal blocks, centrifugal switches, rice cooker sockets).

Compliance Standards: IEC 60695-11-5, GB/T 5169.5, IEC 60884-1, GB 2099.1, VDE 0620, IEC 60335-1, GB 4706.1, UL 60065

Especificaciones del producto

Modelo 6012-HD
Chamber Volume ≥0.5 m³, with glass observation door
Internal Illuminance <20 Lux
Fuel Butane gas (≥95% purity); no air ingress into burner tube (user-provided)
Especificaciones del quemador Length: ≥35mm; Inner diameter: Ø0.5mm±0.1mm; Outer diameter: ≤Ø0.9mm.
Altura de la llama Adjustable: 12mm±1mm (includes measuring gauge).
Temperatura de la llama 100°C±5°C to 700°C±3°C in 23.5±1 seconds. K-type armored thermocouple (Ø0.5mm, imported); optional calibration device included.
Temperature Display Range: 0–1000°C; Resolution: 1°C
Timer Adjustable: 0–99 minutes 99 seconds; Resolution: 0.01s; Configurable ignition/duration recording.
Drive Mechanism Motorized
Key Features Automated control: Burner moves to sample position, shuts off gas post-test, and retracts. 3D-adjustable fixture, integrated ventilation/lighting, and user-friendly operation.
Fuente de alimentación CA 220V, 50Hz