Flame Temperature Calibration Device

Description

Product Description

This calibration device is designed in accordance with IEC 60695-11-5 and GB/T 5169.5 standards for measuring the flame temperature of needle flame test instruments.
During measurement, the copper block is positioned 6±0.5mm above the needle flame burner. The flame is ignited, and the time required for the copper block temperature to rise from 100°C±5°C to 700°C±3°C must be within 23.5±1.0 seconds.

Three measurements are required, with natural cooling of the copper block to below 50°C between each test.

Note: Thermocouples are susceptible to damage above 700°C; thus, the burner must be removed immediately upon reaching this temperature.

Reference Standards
IEC 60695-11-5: Fire hazard testing — Part 11-5: Test flames — Needle flame test method — Apparatus, confirmatory arrangement and guidance.

GB/T 5169.5: Test flames — Needle flame test method — Apparatus, confirmatory test arrangement and guidelines.

Product Specifications

Model 6010-HD
Thermocouple Armored K-type (NiCr/NiAl) fine-wire thermocouple with a 0.5mm diameter sheath.
Timer Adjustable range: 0.01–99.99 seconds.
Copper Block Specifications
Diameter Ø4.00±0.01mm
Length 6.00±0.03mm
Bevel Angle 45 degrees (both ends)
Mass (Pre-drilling) 0.58g±0.01g
Material High-conductivity electrolytic copper
Surface Finish Fully polished